Copyright © 2023 RheoSense, Inc.. All rights reserved. Terms and Conditions & Online Privacy
The next generation m-VROC® II with advanced features and superior repeatability is the ideal viscometer for measuring biological samples such as proteins and mAbs.
The highlighted features are as follows:
A single 15 microliter sample of 262 mg/mL Bovine Gamma Globulin (BγG) in Na phosphate/NaCl buffer was measured over a range of shear rates and at temperatures of 4, 25, and 37 ºC. The testing procedure was as follows:
The (BγG) sample exhibits Newtonian behavior across the range tested, with an average viscosity of 296 cP at 4 ºC, 65 cP at 25 ºC, and 31 cP at 37 ºC. All error bars correspond in length to three times the standard deviation and are smaller than the size of the symbols.